According to IEC60068-2-75 annex B.2,the principle of this calibration procedure is to compare the energy provided by a spring hammer,which is difficult to measure directly,to the energy of a pendulum,calculated from its mass and height of fall.
Et vere malleo ut calibrated est in release basi vere impulsum malleus calibration fabrica et tunc operated ter per a release fabrica; non dimisit manually.
Nam quisque operatio, in striking elementum verticem malleum esse calibrated est in diversis position.The mediocris valorem tria lectiones in calibration machinam est capta est esse ipsam valorem in calibration est in specimine.
Zlt vere impulsum Hammer calibration fabrica quod venundatus est National Institute of Metrology Sina, SCM, grgtest, TSQ, etc.